Z-1176-2017Class II

Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI

Merge Healthcare Incorporated / initiated 2016-04-04 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI on . FDA classified it as Class II and gives the reason as "Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1176-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a letter dated 3/28/2016 via email on 4/4/2016 or via certified mail if they did not have an email address for the customer. A second letter was issued 7/11/2016 to nonresponders.
Quantity in commerce
29 sites have the potentially affected software
Distribution
Distribution was made to medical facilities in AZ, CA, CO, FL, HI, IL, MA, MD, MI, MO, NY, OH, PA, SC, TN, UT, and WI. There was no foreign/government/military distribution.
Product codes and lots
Version 7.0.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1176-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1176-2017Class IITerminatedVoluntary: Firm initiated