Z-1176-2017Class II
Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI
Merge Healthcare Incorporated / initiated 2016-04-04 / Terminated / root cause: software
Merge Healthcare, Inc. began a recall of Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI on . FDA classified it as Class II and gives the reason as "Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1176-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a letter dated 3/28/2016 via email on 4/4/2016 or via certified mail if they did not have an email address for the customer. A second letter was issued 7/11/2016 to nonresponders.
- Quantity in commerce
- 29 sites have the potentially affected software
- Distribution
- Distribution was made to medical facilities in AZ, CA, CO, FL, HI, IL, MA, MD, MI, MO, NY, OH, PA, SC, TN, UT, and WI. There was no foreign/government/military distribution.
- Product codes and lots
- Version 7.0.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1176-2017 | Class II | Terminated | Voluntary: Firm initiated |