510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1192-2017Class II

Integra Cadence Total Ankle System

Integra LifeSciences Corporation / initiated 2017-01-20 / Terminated

Integra LifeSciences Corp. began a recall of Integra Cadence Total Ankle System on . FDA classified it as Class II and gives the reason as "Posterior tibial fractures have been reported". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1192-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra Cadence Total Ankle System
Product code
HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer
Reason for recall
Posterior tibial fractures have been reported.
FDA root cause
Use error
Action
The firm intends to update the surgical technique only. No product will be returned. US and non-US consignees were notified by letter delivered by traceable fax or e-mail on 01/20/2017.
Quantity in commerce
944 units
Distribution
Nationwide
Product codes and lots
Model Number, Item Description, and Lot Number: 10207101, Cadence Tibial Tray, Size 1, Right Lots: QJ0231, QJ0231-G, QJ0467, QJ0467-G; 10207102 Cadence Tibial Tray, Size 2, Right Lots: QJ0227, QJ0227-G, QJ0233, QJ0469, QJ0500, QJ0500-G; 10207103 Cadence Tibial Tray, Size 3, Right Lots: QJ0235, QJ0235-1, QJ0235-1-G, QJ0471, QJ0502-G; 10207104 Cadence Tibial Tray, Size 4, Right Lots: QJ0237, QJ02371, QJ0237-1G, QJ0237-G, QJ0473, QJ0504-G; 10207105 Cadence Tibial Tray, Size 5, Right Lots: QJ0239, QJ0239-1-G, QJ0239G, QJ0475; 10207111 Cadence Tibial Tray, Size 1X, Right Lots: QJ0232, QJ0232-G, QJ0468, QJ0499-G; 10207112 Cadence Tibial Tray, Size 2X, Right Lots: QJ0228, QJ0228-G, QJ0234, QJ0234-G, QJ0470, QJ0501-G; 10207113 Cadence Tibial Tray, Size 3X, Right Lots: QJ0236, QJ0236-1, QJ0236-1-1G, QJ0236-1G, QJ0472, QJ0503, QJ0503-G;  10207114 Cadence Tibial Tray, Size 4X, Right Lots: QJ0238, QJ0238G, QJ0474, QJ0505-G; 10207201 Cadence Tibial Tray, Size 1, Left Lots: QJ0240, QJ0240G, QJ0249, QJ0476; 10207202 Cadence Tibial Tray, Size 2, Left Lots: QJ0242, QJ0242-1-G, QJ0242-2-G, QJ0242-G, QJ0478, QJ0478-G, QJ0507-G;  10207203 Cadence Tibial Tray, Size 3, Left Lots: QJ0244, QJ0244G, QJ0253, QJ0480, QJ0480-G, QJ0509-G; 10207204 Cadence Tibial Tray, Size 4, Left Lots: QJ0246, QJ0246G, QJ0255, QJ0482; 10207205 Cadence Tibial Tray, Size 5, Left Lots: QJ0248, QJ0248G, QJ0257, QJ0484; 10207211 Cadence Tibial Tray, Size 1X, Left Lots: QJ0241, QJ0241G, QJ0477, QJ0477-G, QJ0506;  10207212 Cadence Tibial Tray, Size 2X, Left Lots: QJ0243, QJ0243G, QJ0252, QJ0479, QJ0479-G, QJ0508, QJ0508-G; 10207213 Cadence Tibial Tray, Size 3X, Left Lots: QJ0245, QJ0245G, QJ0481, QJ0481-G, QJ0510, QJ0510-G; 10207214 Cadence Tibial Tray, Size 4X, Left Lots: QJ0247, QJ0247-G, QJ0256, QJ0483, QJ0483-G, QJ0511, QJ0511-G

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1192-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1192-2017Class IITerminatedVoluntary: Firm initiated