The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase...
Siemens Healthcare Diagnostics, Inc. began a recall of The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST)... on . FDA classified it as Class III and gives the reason as "Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1216-2017
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Siemens mailed an Urgent Medical Device Notification letter to all affected customers on May 10, 2016. The notice provided information on actions to be taken by the customer to work around the issue. For further questions, please call (800) 441-9250.
- Quantity in commerce
- 13,147 units
- Distribution
- Distributed throughout the United States
- Product codes and lots
- Lot Number: 4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1216-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1217-2017 | Class III | Terminated | Voluntary: Firm initiated |