Z-1456-2017Class II

Vasoview Hemopro Vh-3500 Endoscopic Vessel Harvesting System

Maquet Cardiovascular LLC / initiated 2017-01-27 / Terminated

Maquet Cardiovascular, LLC began a recall of VASOVIEW HEMOPRO VH-3500 ENDOSCOPIC VESSEL HARVESTING SYSTEM Product Usage: The VASOVIEW HEMOPRO VH-3500 is designed for use in conjunction with the 7-mm... on . FDA classified it as Class II and gives the reason as "Maquet has received several complaints involving the VASOVIEW HEMOPRO VH-3500 which involve the device exhibiting the failure mode of intermittent or no power". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1456-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular LLC
Product
VASOVIEW HEMOPRO VH-3500 ENDOSCOPIC VESSEL HARVESTING SYSTEM Product Usage: The VASOVIEW HEMOPRO VH-3500 is designed for use in conjunction with the 7-mm Endoscope and is intended for specific use with the VASOVIEW HEMOPRO Power Supply.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Maquet has received several complaints involving the VASOVIEW HEMOPRO VH-3500 which involve the device exhibiting the failure mode of intermittent or no power.
FDA root cause
Under Investigation by firm
Action
Maquet Inc. issued an Urgent Removal Correction Notice letter dated January 27, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check inventory for affected device. Remove and keep it in a secure location to preclude being used. Customers were asked to complete the Medical Device Removal Response Form and return to Maquet by e-mailing a scanned copy to [email protected] or by faxing the form to 1-973-629-1518.
Quantity in commerce
4283 units
Distribution
US Nationwide Distribution
Product codes and lots
25126012, 25126080, 25126250, 25126375, 25126542, 25126659, 25128301, 25128602, 25128820, 25128905, 25129016, 25129155, 25129199, 25129239, 25129286, 25129335, 25129416, 25129513

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1456-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1456-2017Class IITerminatedVoluntary: Firm initiated