Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel
Terumo Cardiovascular Systems Corporation began a recall of Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1201-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel
- Product code
- MWS Stabilizer, Heart
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Terumo Cardiovascular Systems has contact all affected parties (user facilities and consignees) via a Customer Notification Letter that advises of the recall activity.
- Quantity in commerce
- The total number of distributed devices stands at 260
- Distribution
- Terumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .
- Product codes and lots
- Serial Number Range 70194 through 70504
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1201-2017 | Class II | Terminated | Voluntary: Firm initiated |