Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment
Terumo Cardiovascular Systems Corporation began a recall of Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1232-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment
- Product code
- MWS Stabilizer, Heart
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.
- FDA root cause
- Other
- Action
- A recall notification is being forwarded to all users and consignees who have received product. The notification of this voluntary recall will be sent to the following consignee in each facility:+ Director of Operating Room Services Proposed Level of Effectiveness Checks: Terumo CVS will verify that 100% of all letters are delivered to affected consignees
- Quantity in commerce
- 1463
- Distribution
- The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.
- Product codes and lots
- Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R Catalog Number 401-231u , Product Description TitanTM 360 Stabilizer Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R, 040 , 048R
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1232-2017 | Class II | Terminated | Voluntary: Firm initiated |