Z-1253-2017Class II

EliA Sample Diluent, Article number, 83-1023-01

Phadia AB / initiated 2017-01-30 / Terminated

Phadia US Inc began a recall of EliA Sample Diluent, Article number, 83-1023-01 on . FDA classified it as Class II and gives the reason as "Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1253-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Phadia AB
Product
EliA Sample Diluent, Article number, 83-1023-01
Product code
NHX Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp)
Reason for recall
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
FDA root cause
Under Investigation by firm
Action
Phadia AB, part of Thermo Fisher Scientific, is voluntarily recalling EliA Sample Diluent, Article number, 83-1023-01, lots JM3Y, JS0B, JP87, JK5N, and JU9C, written notices were mailed via FedEx on 01/30/2017. There have been reports of instrument malfunctions for Phadia 250 due to deformed bottles for the below mentioned product and lots. The deformation of the bottle can cause an erroneous volume detection by the instrument which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples. The deformation is described as a flange or brim, located on the inside lower part of the vial opening. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: Stop using the affected lots of the above mentioned product lots Please, return or attest to scrapping of any unused product lots to the manufacturer and order replacement products free of charge. Review your records to determine if you have had instrument problems due to deformed bottles. If you find the instrument error message 3-145 RIGHT ARM LIQUID DETECTION BELOW LOWER LIMIT Target 2, please collect all log files available and/or contact customer support, who will further assist in assessing possible impact on the test results in scope. Fill in the Medical Device Recall return response on page 3 and return to the manufacturer by e-mail. If customers have any questions, please contact Phadia US, Inc. at 1-800-346- 4364, option #2.
Quantity in commerce
5957
Distribution
Domestic: AR, CA, CO, FL, GA, IL, IN, KY, LA, MI, MN, MO, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, WI, WV Foreign: None VA/DOD: None
Product codes and lots
EliA Sample Diluent, Article number, 83-1023-01, lots JM3Y, JS0B, JP87, JK5N, and JU9C