510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1693-2017Class II

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

Baxter Healthcare Corporation / initiated 2017-01-26 / Terminated / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2 on . FDA classified it as Class II and gives the reason as "SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1693-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2
Product code
FRN Pump, Infusion
Reason for recall
SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
URGENT MEDICAL DEVICE CORRECTION Letters and Reply Forms (dated 1/26/2017) were mailed to customers via USPS first-class mail on 1/26/2017 providing the necessary instructions to customers. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
Quantity in commerce
Domestic: 14,175 units; Canada: 187 units
Distribution
Natiowide, Puerto Rico and Canada
Product codes and lots
14,362 Serial Numbers (please contact CDRH for list of affected Serial Numbers).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1693-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1693-2017Class IITerminatedVoluntary: Firm initiated