Thyroglobulin Antibody (TgAB) Calibrator -
Tosoh Bioscience Inc began a recall of Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin... on . FDA classified it as Class II and gives the reason as "Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1187-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tosoh Bioscience Inc
- Product
- Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.
- Product code
- JZO System, Test, Thyroid Autoantibody
- Reason for recall
- Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 7/24/2012 the firm sent recall notificaiton letters to their customers.
- Quantity in commerce
- 34 boxess/2 sets/6 bottles
- Distribution
- US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.
- Product codes and lots
- Part Number 020391; Lot # B834868, Lot # BX34869, Lot # C234871, Lot # C434872 & Lot # C534873.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1187-2021 | Class II | Terminated | Voluntary: Firm initiated |