Z-1540-2017Class II

Hitachi Oasis MRI System

FUJIFILM Healthcare Americas Corporation / initiated 2015-04-16 / Terminated / root cause: software

Hitachi Medical Systems America Inc began a recall of Hitachi Oasis MRI System on . FDA classified it as Class II and gives the reason as "Image data transferred from the MRI system to a workstation showed errors on the slice position reference image". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1540-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Hitachi Oasis MRI System
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Formal notification letters were sent to consignees on 2/13/17. Hitachi Service will update the software at each site during visits.
Quantity in commerce
210 units
Distribution
US nationwide distribution. The products were distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are four (4) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Building 301 - Andrews Avenue, Fort Rucker, AL 36362; Parma VA Medical Center, 8787 Brookpark Rd, Parma, OH 44129; Viera VA Outpatient Clinic, 2900 Veterans Way, Viera, FL 32940; Perry Point VA Medical Center, Avenue D, Suite 1A-120, Perry Point, MD 21902. There are zero (0) Canada accounts for this recall. There are two (2) Mexico accounts for this recall: Hospital General Tipo "B" de Alta Especialidad - SMH, "Fraccionamiento 1 De La Parcela 101, Del Ejido De Atapaneo", Morelia, MX 58300; Hospital General de Occidente, Zoquipan 1050 Avenue, Zapopan Jalisco, MX 45170 The product was distributed to the following countries: Brazil.
Product codes and lots
M001, M002, M003, M004, M005, M006, M007, M008, M009, M010, M011, M012, M013, M014, M015, M016, M017, M018, M019, M020, M021, M022, M023, M024, M025, M026, M027, M028, M030, M031, M032, M033, M034, M035, M036, M037, M038, M039, M040, M041, M042, M043, M044, M045, M046, M047, M048, M049, M050, M051, M052, M053, M054, M055, M056, M057, M058, M059, M060, M061, M062, M063, M064, M065, M066, M067, M068, M069, M070, M071, M072, M073, M074, M075, M076, M077, M078, M079, M080, M081, M082, M083, M084, M085, M086, M087, M088, M089, M090, M091, M092, M093, M094, M095, M096, M097, M098, M099, M100, M101, M102, M103, M104, M105, M106, M107, M108, M110, M111, M112, M113, M114, M115, M116, M117, M118, M119, M120, M121, M122, M123, M124, M125, M126, M127, M128, M129, M130, M131, M132, M133, M134, M135, M136, M137, M138, M139, M140, M141, M142, M143, M144, M145, M146, M147, M148, M149, M150, M151, M152, M153, M154, M155, M156, M157, M158, M159, M160, M161, M162, M163, M164, M165, M166, M167, M168, M169, M170, M171, M172, M173, M174, M175, M176, M177, M178, M179, M180, M181, M182, M183, M184, M185, M186, M187, M188, M189, M190, M191, M192, M193, M194, M195, M196, M197, M198, M199, M200, M201, M202, M203, M204, M205, M206, M207, M208, M209, M210, M951, M952

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1540-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1540-2017Class IITerminatedVoluntary: Firm initiated
Z-1541-2017Class IITerminatedVoluntary: Firm initiated
Z-1542-2017Class IITerminatedVoluntary: Firm initiated