Z-1220-2017Class II
1.5mm System, Cranio-maxillofacial Bone Screws 1.5 x 4mm HT SD X-DR Screw, Bone Screw
Zimmer GmbH / initiated 2017-01-10 / Terminated
Zimmer Biomet, Inc. began a recall of 1.5mm System, Cranio-maxillofacial Bone Screws 1.5 x 4mm HT SD X-DR Screw, Bone Screw on . FDA classified it as Class II and gives the reason as "The identification of the devices was non-conforming with off-centered slots and the no-go gage starting in the pilot hole". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1220-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 1.5mm System, Cranio-maxillofacial Bone Screws 1.5 x 4mm HT SD X-DR Screw, Bone Screw
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- The identification of the devices was non-conforming with off-centered slots and the no-go gage starting in the pilot hole.
- FDA root cause
- Under Investigation by firm
- Action
- Zimmer Biomet initiated a voluntary recall of 1.5mm System due to identification of non-conformance with off-centered slots and the no-go gage starting in the pilot hole. One distributor with affected product outside of Zimmer Biomet control was notified via email and asked to return the affected product. This communication was sent on via e-mail on January 10, 2017. All product has been returned.There were 89 units in Zimmer Biomet control at the time the removal was initiated, however all 100 are now in Zimmer Biomet control. Customer Service Zimmer Biomet, Inc. P.O. Box 708 Warsaw, IN 46581-0708 Phone: 800-348-2759
- Quantity in commerce
- 11
- Distribution
- Domestic: None VA/DOD: None Foreign: Japan
- Product codes and lots
- 95-6104- Item No. 530430- Lot 1.5x4mm HT SD X-DR Screw