Z-1836-2017Class II

Valleylab Laparoscopic Handset; Material # E2750; Laproscopic Handset Non-Sterile Bulk; Material # E2750NSB

Medtronic, Inc. / initiated 2017-01-26 / Terminated / root cause: design

COVIDIEN MEDTRONIC began a recall of Valleylab Laparoscopic Handset; Material # E2750; Laproscopic Handset Non-Sterile Bulk; Material # E2750NSB Product Usage: For use in 5mm cannulas or larger... on . FDA classified it as Class II and gives the reason as "Customer reports of the device handset continuing to operate after release of the activation button". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1836-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Valleylab Laparoscopic Handset; Material # E2750; Laproscopic Handset Non-Sterile Bulk; Material # E2750NSB Product Usage: For use in 5mm cannulas or larger cannulas with 5mm adapters during laparoscopic and thorascopic procedures. For use with a minimum peak voltage of 3625V.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Customer reports of the device handset continuing to operate after release of the activation button.
Root cause tag
design
FDA root cause
Device Design
Action
A recall was initiated on January 26, 2017. Customers were notified via Federal Express and the letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Disposition of product within Medtronic control and customer returned devices: Will be reworked.
Quantity in commerce
24,703 Units
Distribution
Worldwide - US Nationwide Distribution in the states of: AR AZ CA CO FL GA IA IL KS KY LA MA MI MN MO MS NC NE NM NY OH OR PA SC SD TN TX VA WA WV
Product codes and lots
E2750: 52990142X 52990142X 52990144X 52990144X 52990147X 52990147X 52990151X 52990151X 53250057X 53250057X 53250061X 53250061X 53250063X 53250063X 53250065X 53250065X 53610053X 53610053X 53610055X 53610055X 53610059X 53610059X 53610061X 53610061X 60240055X 60240055X 60240059X 60240059X 60240061X 60240061X 60240065X 60240065X 60520098X 60520098X 60520104X 60520104X 60870064X 60870064X 61220093X 61220093X 61220095X 61220095X 61220098X 61220098X 61300164X 61300164X 61540145X 61540145X 61540150X 61540150X E2750NSB: 52990145X 53250066X 61220096X 61540146X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1836-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1836-2017Class IITerminatedVoluntary: Firm initiated