Z-1235-2017Class II

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single...

Terumo Cardiovascular Systems Corporation / initiated 2015-05-27 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1235-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.
Product code
KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Reason for recall
Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customer Notification Letter/Response Form is being sent to all affected consignees.
Quantity in commerce
579 units have entered into the distribution chain
Distribution
Terumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall
Product codes and lots
TAOS, TC03

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1235-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1235-2017Class IITerminatedVoluntary: Firm initiated