Z-1500-2017Class II

EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs)

Haag-Streit USA Inc / initiated 2017-01-05 / Terminated / root cause: software

Haag-Streit USA Inc began a recall of EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs) on . FDA classified it as Class II and gives the reason as "There is a possibility for data to be stored under the wrong patient on the DICOM Server following a non-standard workflow". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1500-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Haag-Streit USA Inc
Product
EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs).
Product code
NFJ System, Image Management, Ophthalmic
Reason for recall
There is a possibility for data to be stored under the wrong patient on the DICOM Server following a non-standard workflow.
Root cause tag
software
FDA root cause
Software design
Action
On January 5, 2017 the firm sent Field Safety Notice to their customers.
Quantity in commerce
163 units
Distribution
US Nationwide distribution.
Product codes and lots
Model # 7220374 EyeSuite DICOM Option, 7220379 DICOM License .BQ 900 LED S/N: 3682  BQ LET Table S/N: 10038  BX 900 LED w/Digital EyeSuite S/N: 10016  CM 900 S/N: 354 & 491.  IM 900 / BQ 900 LED S/N: 919, 5036, 1136, 3146, 3147, 3148, 3150, 560, 3131, 2412 & 3011.  License Only S/N: 03011, 3210, BP5P922, 5617, 724, 1406, 3204, 2306, 2314, 2514, 3087, 1246, 2395 & 1063.  Lenstar S/N: 167, 285, 332, 416, 469, 656, 878, 1134, 1136, 1197, 1313, 1332, 1607, 2026, 2029, 2030, 2126, 2127, 2128, 2308, 2342, 2343, 2346, 2347, 2349, 2350, 2352, 2353, 2355, 2362, 2387, 2388, 2393, 2686, 2758, 2871, 2901, 2976, 3078, 3084, 3170, 3254, 20213, 3202, 579, 1298, 2055, 2118, 2737, 221, 485, 539, 639, 999, 1246, 2022, 2051, 2274, 2691, 2755, 2756, 2757, 2836, 2908, 3252, 3542, 3753, 3820, 3821, 3901, 3904, 3908, 3946, 3950, 20311, 20111, 20176, 20149 & 10.  Octopus 300 S/N: 4242, 3330, 3735 & 1506.  Octopus 600 S/N: 135, 246, 433, 434, 443, 708, 709, 710, 955, 956, 957, 958 & Demo.  Octopus 900 S/N: 2968, 1784, 2023, 2140, 2237, 2883, 2884, 2886, 2967, 3010, 3286, 3753, 1784, 1286, 1937, 2763, 2778, 3746, 3139, 2813, 1143, 1239, 1287, 2233, 2373, 2406, 2566, 2567, 2780, 2889, 2926, 3200, 2890, 2024, 1582, 1619, 1647, 2174, 2586, 2624, 3006, 3009, 3011 & 3372.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1500-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1500-2017Class IITerminatedVoluntary: Firm initiated