MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305
CareFusion 303, Inc. began a recall of MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305 on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1464-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- An Urgent Medical Device Recall Notification letter will be sent to customers on February 20, 2017 to inform them that CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter model number ME3305 Lot Number 16016790. Customers are informed that CareFusion has received reports of separation and/or leakages between theY-connector and tubing. Customers are informed that leakages from an extension set can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, or underinfusion. The letter informs the customers of the actions to be taken and for recall related questions to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266.
- Quantity in commerce
- 1,200 units
- Distribution
- US Distribution including MA and MD
- Product codes and lots
- Lot No. 16016790
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1464-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.