510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1464-2017Class II

MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305

CareFusion 303, Inc / initiated 2017-02-20 / Terminated / root cause: manufacturing process

CareFusion 303, Inc. began a recall of MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305 on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1464-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305
Product code
FPA Set, Administration, Intravascular
Reason for recall
CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Medical Device Recall Notification letter will be sent to customers on February 20, 2017 to inform them that CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter model number ME3305 Lot Number 16016790. Customers are informed that CareFusion has received reports of separation and/or leakages between theY-connector and tubing. Customers are informed that leakages from an extension set can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, or underinfusion. The letter informs the customers of the actions to be taken and for recall related questions to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266.
Quantity in commerce
1,200 units
Distribution
US Distribution including MA and MD
Product codes and lots
Lot No. 16016790

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1464-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1464-2017Class IITerminatedVoluntary: Firm initiated