510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1432-2017Class II

PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical The intended use of the SIEMENS...

Siemens Medical Solutions USA, Inc. / initiated 2017-01-25 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems... on . FDA classified it as Class II and gives the reason as "Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1432-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens sent a Customer Safety Advisory Notice (CSAN) dated February 7, 2017, to affected customers to inform them of a planned hardware update regarding the mechanics of the overhead suspension of their Digital Linear Accelerator. The notice also explained the actions Siemens is taking to perform the necessary fix at the affected sites. Customers were advised to notify anyone at their organization that should be aware of this information. Also, the safety notice should be forwarded to the new owner and Siemens also requested they be informed of the identity of the device's new owner where possible. Customers with questions were instructed to call 1-800-888-7436. For questions regarding this recall call 610-448-6478.
Quantity in commerce
35 systems
Distribution
Nationwide Distribution
Product codes and lots
Device Model Number 4504200 and 1940035 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1432-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1432-2017Class IITerminatedVoluntary: Firm initiated
Z-1433-2017Class IITerminatedVoluntary: Firm initiated
Z-1434-2017Class IITerminatedVoluntary: Firm initiated
Z-1435-2017Class IITerminatedVoluntary: Firm initiated
Z-1436-2017Class IITerminatedVoluntary: Firm initiated
Z-1437-2017Class IITerminatedVoluntary: Firm initiated