Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous...

Baxter Healthcare Corporation / initiated 2017-02-15 / Terminated / root cause: labeling and UDI

Baxter Healthcare Corporation began a recall of Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing... on . FDA classified it as Class II and gives the reason as "Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1280-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. *Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
Root cause tag
labeling and UDI
FDA root cause
Use error
Action
On February 15, 2017, the firm sent an Urgent Device Correction notification and response form to affected customers via USPS first class mail in order to update software versions of the Prismaflex Control Units. The update was intended to prevent user errors related to the safe unloading of disposable sets. On May 30, 2017, the firm issued an updated notification letter notifying the same customer base of the issue with the programmed syringe sizes. The software update deployed to correct the issue with the safe unloading of disposable sets was also deployed to correct the issue with the programmed syringe sizes. All non-responding customers will be contacted via telephone. Prismaflex control units will be updated to a new software version. For further questions, please call (800) 422-9837.
Quantity in commerce
3,440 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Algeria, Armenia, Australia, Bahrain, Belarus, Belgium, Canada, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, Ireland, Italy, Albania, Bulgaria, Croatia, Czech Republic, Estonia, Hong Kong, Hungary, Iceland, India, Iran, Israel, Ivory Coast, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Montenegro, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Ukraine, Vietnam, Yemen, Argentina, Brazil, Chile, and Dominican Republic
Product codes and lots
Product code: 6023014700, 107493, 113081, 115269; Serial Number: All;   Affected Software Versions: SW 4.1x, 5.10, 6.10, 7.xx, 8.00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1280-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1280-2018Class IITerminatedVoluntary: Firm initiated