Z-1481-2017Class II

882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT...

Philips North America LLC / initiated 2017-01-31 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of 882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar... on . FDA classified it as Class II and gives the reason as "Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1481-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
Root cause tag
software
FDA root cause
Software design
Action
Consignees received a customer information letter via certified mail, 1/31/2017. International consignees were notified by Philips Healthcare Global Market representatives. Field Service Engineers are contacting consignees to conduct appropriate corrections. Non-responders will be contacted multiple times.
Quantity in commerce
1218 total
Distribution
Worldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.
Product codes and lots
11000100, 11000019, 11000034, 11000020, 11000112, 11000086, 11000045, 11000092, 11000037, 11000038, 11000106, 11000059, 11000064, 11000039, 11000087, 11000004, 11000058, 11000030, 11340005, 11000113, 11000117, 11000011, 11000118, 11000074, 11000075, 11000036, 11000104, 11000103, 11000071, 11340002, 11000089, 11000012, 11000041, 11000042, 11000043, 11000049, 11000014, 11000099, 11000022, 11000031, 11000005, 11000028, 11000073, 11000040, 11000077, 11000008, 11000048, 11000093, 11000096, 11000097, 11000069, 11000079, 11000080, 11000024, 11000076, 11000085, 11000108, 11000109, 11000115, 11000056, 11000029, 11340003, 11000044, 11000107, 11000090, 11000017, 11000110, 11000009, 11000035, 11000081, 11000062, 11000013, 11000015, 11000063, 11000083, 11000091, 11000098, 11000054, 11340004, 11000102, 11000061, 11000095, 11000023, 11000094, 11000060, 11000055, 11000070, 11000084, 11000088, 11000068, 11000052, 11000057, 11000050, 11000067, 11000007, 11000010, 11000027, 11000032, 11340001, 11000026, 11000114, 11000078, 11000051, 11000046, 11000003, 11000018, 11000066, 11000047, 11000101, 11000065, 11000072, 11000116, 11000016, 11000105, 11000025, 11000006, 11000033,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1481-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1479-2017Class IITerminatedVoluntary: Firm initiated
Z-1480-2017Class IITerminatedVoluntary: Firm initiated
Z-1481-2017Class IITerminatedVoluntary: Firm initiated