882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT...
Philips Medical Systems (Cleveland) Inc began a recall of 882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar... on . FDA classified it as Class II and gives the reason as "Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1481-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- 882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees received a customer information letter via certified mail, 1/31/2017. International consignees were notified by Philips Healthcare Global Market representatives. Field Service Engineers are contacting consignees to conduct appropriate corrections. Non-responders will be contacted multiple times.
- Quantity in commerce
- 1218 total
- Distribution
- Worldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.
- Product codes and lots
- 11000100, 11000019, 11000034, 11000020, 11000112, 11000086, 11000045, 11000092, 11000037, 11000038, 11000106, 11000059, 11000064, 11000039, 11000087, 11000004, 11000058, 11000030, 11340005, 11000113, 11000117, 11000011, 11000118, 11000074, 11000075, 11000036, 11000104, 11000103, 11000071, 11340002, 11000089, 11000012, 11000041, 11000042, 11000043, 11000049, 11000014, 11000099, 11000022, 11000031, 11000005, 11000028, 11000073, 11000040, 11000077, 11000008, 11000048, 11000093, 11000096, 11000097, 11000069, 11000079, 11000080, 11000024, 11000076, 11000085, 11000108, 11000109, 11000115, 11000056, 11000029, 11340003, 11000044, 11000107, 11000090, 11000017, 11000110, 11000009, 11000035, 11000081, 11000062, 11000013, 11000015, 11000063, 11000083, 11000091, 11000098, 11000054, 11340004, 11000102, 11000061, 11000095, 11000023, 11000094, 11000060, 11000055, 11000070, 11000084, 11000088, 11000068, 11000052, 11000057, 11000050, 11000067, 11000007, 11000010, 11000027, 11000032, 11340001, 11000026, 11000114, 11000078, 11000051, 11000046, 11000003, 11000018, 11000066, 11000047, 11000101, 11000065, 11000072, 11000116, 11000016, 11000105, 11000025, 11000006, 11000033,
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1479-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1480-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1481-2017 | Class II | Terminated | Voluntary: Firm initiated |