Z-1472-2017Class II
BNX Fine Need Aspiration System (19GA); Material/Part #: DSN-19-01
Medtronic, Inc. / initiated 2017-02-08 / Terminated / root cause: manufacturing process
Medtronic began a recall of BNX Fine Need Aspiration System (19GA); Material/Part #: DSN-19-01 on . FDA classified it as Class II and gives the reason as "Handle may break during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1472-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- BNX Fine Need Aspiration System (19GA); Material/Part #: DSN-19-01
- Product code
- FCG Biopsy Needle
- Reason for recall
- Handle may break during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- A recall was initiated on February 8, 2017. Customers were notified via Federal Express and the letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 1856 Corporate Dr. Suite 140 Norcross, GA 30093, Attention: Field Returns Department. Disposition of product within Medtronic control and customer returned devices: Will be reworked.
- Quantity in commerce
- 1,096 Units (total for all products)
- Distribution
- US Distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI; and countries of DK and DE
- Product codes and lots
- Lot Number: B000000176, B000000181
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1472-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1473-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1474-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1475-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1476-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.