EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the...
Becton Dickinson & Co. began a recall of EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The... on . FDA classified it as Class II and gives the reason as "The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1414-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Customers were initially contacted via phone calls and subsequently provided with a written communication. For questions contact BD Technical Services Department at 1-800-638-8663.
- Quantity in commerce
- Qty Distr. to Field - 115
- Distribution
- 15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
- Product codes and lots
- Lot Numbers Version 6.20A Cat. Number 442953
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1414-2017 | Class II | Terminated | Voluntary: Firm initiated |