510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1414-2017Class II

EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the...

Becton Dickinson & Co / initiated 2014-08-25 / Terminated / root cause: supplier

Becton Dickinson & Co. began a recall of EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The... on . FDA classified it as Class II and gives the reason as "The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1414-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Customers were initially contacted via phone calls and subsequently provided with a written communication. For questions contact BD Technical Services Department at 1-800-638-8663.
Quantity in commerce
Qty Distr. to Field - 115
Distribution
15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
Product codes and lots
Lot Numbers Version 6.20A  Cat. Number 442953

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1414-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1414-2017Class IITerminatedVoluntary: Firm initiated