Z-1406-2017Class II
A-series lift - A625 Manual Part No. 324117. Prism Medical
Handicare USA, Inc. / initiated 2017-01-30 / Terminated / root cause: design
ErgoSafe Products, LLC (DBA) Prism Medical began a recall of A-series lift - A625 Manual Part No. 324117. Prism Medical on . FDA classified it as Class II and gives the reason as "During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1406-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Handicare USA, Inc.
- Product
- A-series lift - A625 Manual Part No. 324117. Prism Medical.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm called each customer on Jan 30, 2017 to review the recall and schedule a time for the firm to come to their facility to personally swap the recalled product for a new device. The FSNs were sent on Jan 31, 2017.
- Quantity in commerce
- 76
- Distribution
- Product was distributed to MT and NV; one VA consignee.
- Product codes and lots
- A6MC000016 A6MC000019 A6MC000020 A6MC000021 A6MC000022 A6MC000023 A6MC000024 A6MC000025 A6MC000026 A6MC000027 A6MC000028 A6MC000029 A6MC000030 A6MC000031 A6MC000009 A6MC000002 A6MC000004 A6MC000001 A6MC000003 A6MC000005 A6MC000006 A6MC000007 A6MC000008 A6MC000010 A6MC000011 A6MC000012 A6MC000013 A6MC000014 A6MC000015 A6MC000017 A6MC000018 A6MC000096 A6MC000087 A6MC000088 A6MC000090 A6MC000091 A6MC000092 A6MC000093 A6MC000094 A6MC000048 A6MC000068 A6MC000070 A6MC000072 A6MC000073 A6MC000074 A6MC000075 A6MC000078 A6MC000079 A6MC000081 A6MC000082 A6MC000083 A6MC000085 A6MC000086 A6MC000042 A6MC000043 A6MC000045 A6MC000046 A6MC000047 A6MC000049 A6MC000050 A6MC000051 A6MC000052 A6MC000053 A6MC000054 A6MC000055 A6MC000056 A6MC000057 A6MC000058 A6MC000059 A6MC000060 A6MC000061 A6MC000062 A6MC000063 A6MC000064 A6MC000066 A6MC000071
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1405-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1406-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.