Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets...

Datascope Corp. / initiated 2017-02-16 / Terminated / root cause: design

Datascope Corporation began a recall of Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They... on . FDA classified it as Class II and gives the reason as "A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1769-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
Product code
DWE Tubing, Pump, Cardiopulmonary Bypass
Reason for recall
A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet Inc. was notified by their customer on January 1, 2017. This was a reverse notification. The one impacted customer said they will not return the product since they can easily correct the pigtail connection before use. For further questions regarding this recall, please call (973) 709-7442.
Quantity in commerce
21 kits
Distribution
US Distribution to the state of : CA.
Product codes and lots
Lot number 3000033848

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1769-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1769-2017Class IITerminatedVoluntary: Firm initiated