Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000...
ConMed Corporation began a recall of Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 100 VOLTS AC... on . FDA classified it as Class II and gives the reason as "For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1756-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Recalling firm
- ConMed Corporation
- Product
- Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 100 VOLTS AC, REF/Catalog Number 7-900-100, Rx Only. There is no problem with the Hyfrecator 2000 Electrosurgical Unit. -- ASSEMBLED IN MEXICO --- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- ConMed sent an URGENT: MEDICAL DEVICE RECALL Notices (dated 2/28/2017)to all affected US customers by USPS First Class Mail. International letters are in the process of translation and are scheduled to be sent by March 15, 2017. Customers were instructed to review their inventory for any of the affected product and were asked to complete the business reply form and return it to ConMed Corporation, 525 French Road, Utica, New York 13502. Customers with questions or requests should contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email [email protected]. For questions regarding this recall call 315-624-3533.
- Distribution
- Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
- Product codes and lots
- Serial Numbers: 1540139, 15280024, 15420101, 15470153, 15510179, 16010007, 1540144, 15280026, 15420102, 15470161, 15510182, 16010008, 15250002, 15290032, 15430106, 15490169, 15510186, 16010009, 15250004, 15290033, 15430107, 15490170, 15510189, 16010010, 15250005, 15300038, 15430110, 15490172, 15510190, 16430108, 15250006, 15350041, 15430111, 15500173, 15510192, 16430109, 15250007, 15350043, 15430113, 15500174, 16010001, 16430112, 15250008, 15350044, 15430114, 15500175, 16010002, 15260010, 15360054, 15430115, 15500176, 16010003, 15260012, 15360057, 15460140, 15500177, 16010004, 15260013, 15360059, 15460142, 15510178, 16010005, 15280023, 15360060, 15460146, 15510179, 16010006.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1756-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1757-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1758-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1759-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1760-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.