Z-1469-2017Class II
OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
GE Healthcare / initiated 2017-02-21 / Terminated / root cause: manufacturing process
GE OEC Medical Systems, Inc began a recall of OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems on . FDA classified it as Class II and gives the reason as "OEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1469-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- OEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On February 21, 2017, an Urgent Medical Device Correction letter was sent to one consignee with an affected unit; a field service representative hand delivered the customer communication.
- Quantity in commerce
- 1
- Distribution
- One consignee in CA.
- Product codes and lots
- OEC FlexiView 8800 systems that had NEMA 5-15 15 Amp power cords replaced between January 28, 2016 and November 23, 2016.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1469-2017 | Class II | Terminated | Voluntary: Firm initiated |