Z-0318-2018Class II

HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606...

Repro-Med Systems, Inc. / initiated 2016-03-10 / Terminated / root cause: manufacturing process

Repro-Med Systems, Inc. began a recall of HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609... on . FDA classified it as Class II and gives the reason as "On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0318-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Repro-Med Systems, Inc.
Product
HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.
Product code
FPA Set, Administration, Intravascular
Reason for recall
On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected parties. Customers are encouraged to complete and return the acknowledgement form via fax to: 845-469-5518.
Distribution
Nationwide in US
Product codes and lots
Lot Numbers: 7.069/16, 7.068/16, 7.066/16, 7.065/16, 7.064/16, 7.063/16, 7.062/16, 7.061/16, 7.060/16, 7.059/16, 7.058/16, 7.057/16, 7.056/16, 7.055/16, 7.054/16, 7.053/16, 7.052/16, 7.051/16, 7.050/16, 7.049/16, 7.048/16 & 7.046/16.  Expiration: 2019-02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0318-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0317-2018Class IITerminatedVoluntary: Firm initiated
Z-0318-2018Class IITerminatedVoluntary: Firm initiated