510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1760-2017Class II

HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED...

ConMed Corporation / initiated 2017-02-28 / Terminated / root cause: labeling and UDI

ConMed Corporation began a recall of HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. ---... on . FDA classified it as Class II and gives the reason as "For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1760-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Recalling firm
ConMed Corporation
Product
HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
ConMed sent an URGENT: MEDICAL DEVICE RECALL Notices (dated 2/28/2017)to all affected US customers by USPS First Class Mail. International letters are in the process of translation and are scheduled to be sent by March 15, 2017. Customers were instructed to review their inventory for any of the affected product and were asked to complete the business reply form and return it to ConMed Corporation, 525 French Road, Utica, New York 13502. Customers with questions or requests should contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email [email protected]. For questions regarding this recall call 315-624-3533.
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
Product codes and lots
Affected Lot Numbers: 201511234, 201512074, 201512094, 201601064, 201603284

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1760-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1756-2017Class IITerminatedVoluntary: Firm initiated
Z-1757-2017Class IITerminatedVoluntary: Firm initiated
Z-1758-2017Class IITerminatedVoluntary: Firm initiated
Z-1759-2017Class IITerminatedVoluntary: Firm initiated
Z-1760-2017Class IITerminatedVoluntary: Firm initiated