Z-1536-2017Class II

9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable...

Medical Components, Inc. / initiated 2017-02-23 / Terminated / root cause: labeling and UDI

Medical Components, Inc dba MedComp began a recall of 9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies... on . FDA classified it as Class II and gives the reason as "The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1536-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medical Components, Inc.
Product
9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer during the insertion procedure.
Root cause tag
labeling and UDI
FDA root cause
Unknown/Undetermined by firm
Action
Medcomp mailed a Customer Notification letter dated February 23, 2017, to all affected customers to inform them of the issue. Customers were asked to examine their inventory and return all unused product. Distributors were also asked to locate the end user/facility of where the affected product was sold and communicate the recall to them. Customers were instructed to contact their customer service representative for a Returned Goods Authorization (RGA) number if necessary at 215-256-4201. Customers were also instructed to complete and return page 2 of the letter by fax to 215-256-9191. For questions regarding this recall call 215-256-4201.
Quantity in commerce
71 units
Distribution
Worldwide Distribution to PR and Panama
Product codes and lots
MRCTI96801 Lot# MHWQ060 S2 , UDI# 884908031232

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1536-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1536-2017Class IITerminatedVoluntary: Firm initiated