Cobas 8100 uni-directional reformatter (BRF) module with Software Version 02-xx
Roche Diagnostics Corporation began a recall of Cobas 8100 uni-directional reformatter (BRF) module with Software Version 02-xx on . FDA classified it as Class II and gives the reason as "Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (BRF) modules with software version 02-xx, only...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1764-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Cobas 8100 uni-directional reformatter (BRF) module with Software Version 02-xx
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (BRF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- Roche initiated a recall on 03/01/2017, notices were issued via UPS to customers and instructed customers to do the following: Roche is developing software version 03-01 to resolve this issue, and a Roche Representative will be contacting you in the near future to schedule an update. Until this software update is available, the rack buffering option for the BRF and URF modules must be disabled. If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined below to disable rack buffering. If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. If you have a major spill, set the affected module to offline and contact the Roche Support Network Customer Support Center at 1-800-428-2336. " If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined in this Urgent Medical Device Correction (UMDC) to disable rack buffering. " If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. " Complete the attached faxback form and fax it to 1-888-345-0024. " File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Quantity in commerce
- 45
- Distribution
- Domestic: MA, IN, AL, OH, MO, NE, LA, CA, IA, NJ, AR, PA, SC, IL, TX, and MI. Foreign: None
- Product codes and lots
- Cobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xx Catalog Number: URF module: 07441657001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1763-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1764-2017 | Class II | Terminated | Voluntary: Firm initiated |