Z-1478-2017Class II
Inion GRT Tack, drill bit and tack appl., 4x, Sterile. Is indicated for the surgical treatment of periodontal...
Inion Oy / initiated 2017-03-01 / Terminated / root cause: packaging
Inion Ltd. began a recall of Inion GRT Tack, drill bit and tack appl., 4x, Sterile. Is indicated for the surgical treatment of periodontal defects to aid in the regeneration and... on . FDA classified it as Class II and gives the reason as "Aluminum pouch seal was noticed to be defective (slightly open)". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1478-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Inion Oy
- Product
- Inion GRT Tack, drill bit and tack appl., 4x, Sterile. Is indicated for the surgical treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures, for pre-implant and peri-implant surgery and for covering bone defects and empty sockets.
- Product code
- NPK Barrier, Synthetic, Intraoral
- Reason for recall
- Aluminum pouch seal was noticed to be defective (slightly open).
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Inion sent an Urgent: Medical Device Recall Notification letter dated March 1, 2017. The recall letter asked customers to inspect the aluminum pouch seals, remove any devices with defective seals from sales stock, and to return the product to Inion. The customers were also asked to inform Inion Oy of the outcome of their inspections using a feedback form. For further questions call (954) 659-9224.
- Quantity in commerce
- 36 pieces
- Distribution
- US Distribution to : FL and NC.
- Product codes and lots
- Lot Number 1604006, 1505007, 1503039
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|