Z-1710-2017Class II

Access 2 assay Immunoassay System, Remanufactured, Part No. 386220

Siemens Healthcare Diagnostics Inc. / initiated 2017-03-07 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1710-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access 2 assay Immunoassay System, Remanufactured, Part No. 386220 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.
Root cause tag
software
FDA root cause
Software Design Change
Action
An Urgent Medical Device Recall letter was sent on 3/7/17 to inform customers that Beckman Coulter has determined through internal investigation that a circuit board that controls motor functions in the instrument systems listed above can affect device positioning on the pipettor and incubator belt while the instrument is running. Customers are informed of the impact, action to be taken, and the resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
Quantity in commerce
851 units total for all products (291 units in US)
Distribution
Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe
Product codes and lots
Serial No. 501012,501110,501190,501224,501294,501371,501394,501403,501440, 501443,501490,501501,501673,501681,501693,501861,501920,501926, 501936,501970,502165,502196,502319,502331,502332,502368,502458, 502475,502506,502543,502549,502556,502574,502640,502647,502651, 502791,502843,502848,503153,503279,503313,503354,503384,503392, 503424,503543,503627,503678,504046,504071,504072,504092,504112, 504195,504322,504365,504370,504756,504767,504902,505042,505195, 505251,505254,505644,505795,505840,505947,506002,506026,506108, 506119,506153,506187,506371,506406,506560,506586,506664,506708, 506766,506786,506837,506846,507039,507120,507161,507162,507214, 507229,507248,507282,507353,507397,507480,507511,507604,507619, 507842,507929,508019,508096,508179,508227,508313,508334,508412, 508478,508509,508516,508767,509051,509087,509113,509164,509253, 509297,509632,509837,510049,511042,511047,511180,511383,700024, 800321,800341,800443,800608,800611,800650,800849

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1710-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1709-2017Class IITerminatedVoluntary: Firm initiated
Z-1710-2017Class IITerminatedVoluntary: Firm initiated
Z-1711-2017Class IITerminatedVoluntary: Firm initiated