Fem IM Nail 15mmdx46cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail...
Zimmer Biomet, Inc. began a recall of Fem IM Nail 15mmdx46cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants on . FDA classified it as Class II and gives the reason as "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2555-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Fem IM Nail 15mmdx46cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Zimmer Biomet initiated a field action for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures. Zimmer Biomet sent letters via Fed-Ex and e-mail dated 02/28/2017 to customers informing them of the recall, and instructing them to do the following: 1. Review this notification and ensure affected team members are aware of the contents. 2. Complete the Certification of Acknowledgement portion of Attachment 1 a. Return a digital copy to [email protected] within three (3) days. 3. Assist the Zimmer Biomet sales representative quarantine all affected product. 4. If after reviewing this notice customers have further questions or concerns, please call 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an oncall representative in the case of an emergency. Alternatively, questions may be sent by email to [email protected]. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: MedWatch Reporting: Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by mail, or by fax. Online: www.fda.gov/medwatch/report.htm Mail: Use postage paid, pre-addressed form FDA 3500, available at: www.fda.gov/MedWatch/getforms.htm Fax: 1-800-FDA-0178 Under 21 CFR 803, manufacturers are required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Zimmer Biomet informed of any adverse events associated with this product or any other Zimmer Biomet product by emailing [email protected]. For questions regarding this recall call 574-371-3071.
- Distribution
- Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, Thailand
- Product codes and lots
- Item Number:00225246015, Lot Number Expiry Date Before: March 2027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2549-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2550-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2551-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2552-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2553-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2554-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2555-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2556-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2557-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2558-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2559-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2560-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2561-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2562-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2563-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2564-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2565-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2566-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2567-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2568-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2569-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2570-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2571-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2572-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2573-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2574-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2575-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2576-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2577-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2578-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2579-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2580-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2581-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2582-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2583-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.