510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1515-2017Class II

Poly Component Trial, CR, packaged in the following sizes and configurations: a) Poly Component Trial - CR...

Medtronic Sofamor Danek USA Inc / initiated 2017-02-20 / Terminated

Medtronic Sofamor Danek USA Inc began a recall of Poly Component Trial, CR, packaged in the following sizes and configurations: a) Poly Component Trial - CR 3x10, REF 90-SRK-160310 b) Poly Component Trial - CR... on . FDA classified it as Class II and gives the reason as "The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1515-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
Poly Component Trial, CR, packaged in the following sizes and configurations: a) Poly Component Trial - CR 3x10, REF 90-SRK-160310 b) Poly Component Trial - CR 3x12, REF 90-SRK-160312 c) Poly Component Trial - CR 3x14, REF 90-SRK-160314 d) Poly Component Trial - CR 4x10, REF 90-SRK-160410 e) Poly Component Trial - CR 4x12, REF 90-SRK-160412 f) Poly Component Trial - CR 4x14, REF 90-SRK-160414 g) Poly Component Trial - CR 5x10,REF 90-SRK-160510 h) Poly Component Trial - CR 5x12, REF 90-SRK-160512 i) Poly Component Trial - CR 5x14, REF 90-SRK-160514 j) Poly Component Trial - CR 6x10, REF 90-SRK-160610 k) Poly Component Trial - CR 6x12, REF 90-SRK-160612 l) Poly Component Trial - CR 6x14, REF 90-SRK-160614 The Responsive Orthopedics (RO) Total Knee Arthroplasty (TKA) System is intended to restore alignment, stability, range of motion, and alleviate pain by replacing the articulating surfaces of the knee joint
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.
FDA root cause
Under Investigation by firm
Action
Medtronic initiated their recall on 20 February 2017, by letter delivered by Fed Ex . The letter was addressed to the Risk/Materials Manager and stated that a Medtronic Clinical Specialist would coordinate the removal of the Responsive Orthopedics Total Knee Arthroplasty (TKA) System instrument sets from the premises. Customers with questions were instructed to contact the Medtronic Post-Market Quality Team at 901-399-3197.
Distribution
US distribution to AR, MA
Product codes and lots
Lot Numbers: a) REF 90-SRK-160310, TU53730-01 b) REF 90-SRK-160312, TU53731-01 c) REF 90-SRK-160314, TU53732-01 d) REF 90-SRK-160410, TU53730-02 e) REF 90-SRK-160412, TU53731-02 f) REF 90-SRK-160414, TU53732-02 g) REF 90-SRK-160510, TU53730-03 h) REF 90-SRK-160512, TU53731-03 i) REF 90-SRK-160514, TU53732-03 j) REF 90-SRK-160610, TU53730-04 k) REF 90-SRK-160612, TU53731-04 l) REF 90-SRK-160614, TU53732-04

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1515-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1513-2017Class IITerminatedVoluntary: Firm initiated
Z-1514-2017Class IITerminatedVoluntary: Firm initiated
Z-1515-2017Class IITerminatedVoluntary: Firm initiated
Z-1516-2017Class IITerminatedVoluntary: Firm initiated