Z-1496-2017Class II

Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare...

Merge Healthcare Incorporated / initiated 2015-09-22 / Terminated / root cause: manufacturing process

Merge Healthcare, Inc. began a recall of Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI on . FDA classified it as Class II and gives the reason as "For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1496-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016, via email the week of April 4, 2016, or via certified mail if they did not have an email address for the customer. On/about September 22, 2015, the recalling firm emailed "IMPORTANT PRODUCT INFORMATION" letters dated August 20, 2015, to their customers. The letter notifies the customer of the issue and informs them a fix for the issue is in the new software release. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. The response was required no later than April 18, 2016. Nonresponders were issued a second letter dated July 11, 2016, asking for a response by August 2, 2016. The letter and response form contained the same information as the first letter. If the second letter results in additional unresponsive customers, a call campaign will be managed by Customer Support and documented via SalesForce. The recalling firm reported emails will be tracked for delivery confirmation and the letters will be tracked for delivery by USPS certified services. For further questions, please call (877) 741-5369. .
Quantity in commerce
16 sites were originally identified as potentially having the affected versions
Distribution
US Distribution was made to medical facilities located in GA, IL, MD, MI, MO, OH, TN, and VT. There was no foreign/government/military distribution.
Product codes and lots
Versions 8.30, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 10.0, 10.0.1, and 10.1 with Issuer of Patient ID (IPID)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1496-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1496-2017Class IITerminatedVoluntary: Firm initiated