Z-1683-2017Class II
Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For...
National Biological Corp. / initiated 2017-01-27 / Terminated / root cause: manufacturing process
National Biological Corp began a recall of Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment on . FDA classified it as Class II and gives the reason as "The wiring may allow the lamps to turn on with the key switch rather than the timer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1683-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- National Biological Corp.
- Product
- Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment.
- Product code
- FTC Light, Ultraviolet, Dermatological
- Reason for recall
- The wiring may allow the lamps to turn on with the key switch rather than the timer.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Individual patients were contacted via telephone. Replacement devices have been distributed. The distributor was contacted via email. The firm has requested that consignees return the recalled device to their firm to be held under quarantine.
- Quantity in commerce
- 7 units
- Distribution
- Worldwide distribution. US states: IA, IL, MI, NY, PA, and WI. Country: Jordan. There are no government accounts for this recall. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
- Product codes and lots
- Serial numbers: 39870, 39873, 39874, 39866, 39867, 39871, 39869