Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1683-2017Class II

Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For...

National Biological Corp. / initiated 2017-01-27 / Terminated / root cause: manufacturing process

National Biological Corp began a recall of Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment on . FDA classified it as Class II and gives the reason as "The wiring may allow the lamps to turn on with the key switch rather than the timer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1683-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
National Biological Corp.
Product
Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment.
Product code
FTC Light, Ultraviolet, Dermatological
Reason for recall
The wiring may allow the lamps to turn on with the key switch rather than the timer.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Individual patients were contacted via telephone. Replacement devices have been distributed. The distributor was contacted via email. The firm has requested that consignees return the recalled device to their firm to be held under quarantine.
Quantity in commerce
7 units
Distribution
Worldwide distribution. US states: IA, IL, MI, NY, PA, and WI. Country: Jordan. There are no government accounts for this recall. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
Product codes and lots
Serial numbers: 39870, 39873, 39874, 39866, 39867, 39871, 39869

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1683-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1683-2017Class IITerminatedVoluntary: Firm initiated