Z-1801-2017Class II

VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the...

bioMerieux, Inc. / initiated 2017-03-17 / Terminated / root cause: design

BioMerieux SA began a recall of VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK... on . FDA classified it as Class II and gives the reason as "Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1801-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).
Product code
CHP Radioimmunoassay, Estradiol
Reason for recall
Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.
Root cause tag
design
FDA root cause
Device Design
Action
bioMerieux sent an Important Product Safety Notice A Product Safety Notice to all affected customers on March accounts on March 17, 2017, informing them not to use VIDAS Estradiol II test to test estradiol level in patients under Fulvestrant therapy. The letter did not request return of any of the product, and requested that they return an acknowledgement form acknowledging receipt of the notice and that they have followed the instructions and implemented the actions in the field notice. Customers with questions were instructed to contact their bioMerieux Customers Service representative. For questions regarding this recall call 314-731-8805.
Quantity in commerce
834 kits
Distribution
Nationwide Distribution including AZ, CA, CO, FL, GA, IN, IA, LA, MI, MS, NY, NC, OH, OK, OR, PA, TX, WA
Product codes and lots
Lots 1005081700, 1005172850 & 1005359170

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1801-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1801-2017Class IITerminatedVoluntary: Firm initiated