Z-1823-2017Class II
Tapered SwissPlus Implant System, Model No. SPB10
ZimVie US Corp LLC / initiated 2017-03-20 / Terminated / root cause: manufacturing process
Zimmer Dental Inc began a recall of Tapered SwissPlus Implant System, Model No. SPB10 on . FDA classified it as Class II and gives the reason as "It was manufactured with the minor diameter of the internal thread that is below specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1823-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZimVie US Corp LLC
- Product
- Tapered SwissPlus Implant System, Model No. SPB10
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- It was manufactured with the minor diameter of the internal thread that is below specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- A customer letter was sent on 3/20/17 to inform customers that Zimmer Biomet is recalling a single lot of the Tapered SwissPlus 3.7mmD Implant, Lot No. 63498673. The letter informs customers that the recall is due to one (1) lot of part number SPB10 (lot 63498673) that was manufactured with the minor diameter of the internal thread below specification. This condition may cause mating issues with some screws. The condition occurs when the screw used has a minor thread diameter that is closer to the maximum allowed diameter (maximum material condition). However, screws closer to the least material condition will still fit into the implant. Customers with questions or concerns are instructed to call the customer call center at 1-800-854-7091 between 8:00 am and 5:00pm EST, Monday through Friday. Customers are also instructed to email to [email protected].
- Quantity in commerce
- 188 units
- Distribution
- Worldwide: ZA, FR, IT, IR
- Product codes and lots
- Lot #63498673; UDI: (01) 008890240177 (10) 63498673 (17) 200731
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1823-2017 | Class II | Terminated | Voluntary: Firm initiated |