510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1823-2017Class II

Tapered SwissPlus Implant System, Model No. SPB10

ZimVie US Corp LLC / initiated 2017-03-20 / Terminated / root cause: manufacturing process

Zimmer Dental Inc began a recall of Tapered SwissPlus Implant System, Model No. SPB10 on . FDA classified it as Class II and gives the reason as "It was manufactured with the minor diameter of the internal thread that is below specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1823-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Tapered SwissPlus Implant System, Model No. SPB10
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
It was manufactured with the minor diameter of the internal thread that is below specification.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A customer letter was sent on 3/20/17 to inform customers that Zimmer Biomet is recalling a single lot of the Tapered SwissPlus 3.7mmD Implant, Lot No. 63498673. The letter informs customers that the recall is due to one (1) lot of part number SPB10 (lot 63498673) that was manufactured with the minor diameter of the internal thread below specification. This condition may cause mating issues with some screws. The condition occurs when the screw used has a minor thread diameter that is closer to the maximum allowed diameter (maximum material condition). However, screws closer to the least material condition will still fit into the implant. Customers with questions or concerns are instructed to call the customer call center at 1-800-854-7091 between 8:00 am and 5:00pm EST, Monday through Friday. Customers are also instructed to email to [email protected].
Quantity in commerce
188 units
Distribution
Worldwide: ZA, FR, IT, IR
Product codes and lots
Lot #63498673; UDI: (01) 008890240177 (10) 63498673 (17) 200731

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1823-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1823-2017Class IITerminatedVoluntary: Firm initiated