Z-1767-2017Class II

CADD Administration Set, 21-7339-24, Rx Only, Sterile EO

Abbott Medical / initiated 2017-03-17 / Terminated / root cause: packaging

Smiths Medical ASD Inc. began a recall of CADD Administration Set, 21-7339-24, Rx Only, Sterile EO on . FDA classified it as Class II and gives the reason as "Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1767-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
CADD Administration Set, 21-7339-24, Rx Only, Sterile EO.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on the IFU correctly identifies the product to include a 0.2 micron filter, the illustration on the IFU identifies the filter size as 2.0 microns.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Smiths Medical sent an "Urgent Medical Device Field Safety Notice" dated March 13, 2017. The letter was addressed to "Risk Manager". It provided a list of the Affected devices, Risk to Health, Instructions to Distributors & Customers. Advised consignees to notify their customers. There is no need to return the devices to Smiths Medical. Requested consignees to complete the Urgent Medical Device Field Safety Notice (FSN) Response Form and return to [email protected]. For questions contact Stericycle at 844-265-7407. For further questions regarding this recall , please call (844) 265-7407.
Quantity in commerce
38,415 (38145 US; 2268 OUS)
Distribution
US:AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WV, WY. OUS: Canada
Product codes and lots
46X817, 46X829, 46X899, 46X828, 46X896, 46X990, 46X831, 46X897, 46X1001.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1767-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1767-2017Class IITerminatedVoluntary: Firm initiated