Z-1748-2017Class II

Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 The...

Siemens Healthcare Diagnostics Inc. / initiated 2017-03-08 / Terminated

Siemens Healthcare Diagnostics, Inc began a recall of Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 The ADVIA Centaur BR assay is an in vitro diagnostic... on . FDA classified it as Class II and gives the reason as "Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1748-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur, ADVIA Centaur XP, and ADVIA Centaur XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.
Product code
MOI System, Test, Immunological, Antigen, Tumor
Reason for recall
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
FDA root cause
Under Investigation by firm
Action
Siemens Healthcare issued An Urgent Medical Device Correction (UMDC) to customers in the United States on March 08, 2017 to be delivered to customers on March 09, 2017and (UFSN) was sent for distribution to customers outside the United States on March 08, 2017. These notices inform customers of the potential and mitigation action to be taken for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) assay on the ADVIA Centaur Systems. This issue affects all current and future lots of ADVIA Centaur Systems BR reagents until a solution is implemented. Fax-back forms included with letter used to confirm that the customers received the Urgent Medical Device Correction to the Customer Care Center at 312-275-7795. For questions regarding this recall call 508-668-5000. .
Quantity in commerce
1900 kits
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Australia, Brazil, Canada, Columbia, Cyprus, Japan, Spain
Product codes and lots
All lot codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1748-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1747-2017Class IITerminatedVoluntary: Firm initiated
Z-1748-2017Class IITerminatedVoluntary: Firm initiated
Z-1749-2017Class IITerminatedVoluntary: Firm initiated