Z-2166-2017Class II

SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems...

Siemens Healthcare Diagnostics Inc. / initiated 2017-01-31 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in... on . FDA classified it as Class II and gives the reason as "Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8C, as published in the Instructions for Use (IFU). Adjustors...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2166-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Product code
CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone
Reason for recall
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens issued an Urgent Medical Device Correction (UMDC) to customers using the affected SHBG adjustor lots. Customers were advised to store the affected reconstituted adjustor lots at 2-8C or -20C for up to 14 days, after which they must be discarded. The IMMULITE/IMMULITE 1000 SHBG is not affected by this issue. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 914-631-8000.
Quantity in commerce
9100 kits
Distribution
Nationwide Distribution
Product codes and lots
Lot Numbers: 1) 357, 2) 358, 3) 360, 4) 361 & 5) 363  Expiration Dates: 1) 2017-01-30, 2) 2017-03-31, 3) 2017-04-30, 4) 2017-05-31 & 5) 2017-06-30

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2166-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2166-2017Class IITerminatedVoluntary: Firm initiated