Boston Scientific Imager II Angiographic Catheter, a pathway for delivering contrast media to selected sites...
Boston Scientific Corporation began a recall of Boston Scientific Imager II Angiographic Catheter, a pathway for delivering contrast media to selected sites in the vascular system. M001314141 (5-pack outer... on . FDA classified it as Class II and gives the reason as "A field report indicated some units within this lot were missing the bottom pouch seal. The compromised seal is completely missing and is obvious to the user". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2645-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Imager II Angiographic Catheter, a pathway for delivering contrast media to selected sites in the vascular system. M001314141 (5-pack outer package UPN), M001314140, single unit, inner package UPN)
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- A field report indicated some units within this lot were missing the bottom pouch seal. The compromised seal is completely missing and is obvious to the user.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Customers were sent an Urgent Medical Device Removal letter, dated 12/15/2016, via overnight mail delivery beginning 12/15/2016. The letter identified affected product, described the issue, and asked for product to be returned. Removal instructions were provided and a Reply Verification Tracking Form was asked to be returned. Questions can be directed to Nicole Pshon, Field Action Manager at 763-494-1133 or [email protected]
- Quantity in commerce
- 39 boxes of 5 (195 units)
- Distribution
- CA, FL, KY, MO, NM, OK, OR, PA, and TX.
- Product codes and lots
- Lot: 00112758, Use by 2018-05-11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2645-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY