Z-1832-2017Class I

HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit

HeartWare, Inc. / initiated 2017-03-10 / Open, Classified / root cause: design

Heartware began a recall of HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been... on . FDA classified it as Class I and gives the reason as "In April 2015, HeartWare, now a part of Medtronic, notified users of a safety issue related to the HVAD Pump Driveline Splice Kit. Medtronic now intends to send a follow-up communication to identified clinicians...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1832-2017
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartWare, Inc.
Product
HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been identified. These repairs are limited to the portion of the driveline that is outside the patient boundary (exit side).
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
In April 2015, HeartWare, now a part of Medtronic, notified users of a safety issue related to the HVAD Pump Driveline Splice Kit. Medtronic now intends to send a follow-up communication to identified clinicians informing them that a new, improved splice kit is available.
Root cause tag
design
FDA root cause
Device Design
Action
HeartWare, now a part of Medtronic sent an Urgent Medical Device Recall letter on April 2015 and a follow-up letter on March 2017, to customers via overnight mail and/or email.. The letters identified the affected product, problem and actions to be taken. Customers were instructed to review the recall notice, forward the notice to individuals within their organization who need to be aware of the notice, complete, sign, and return the Acknowledgement Form . For questions contact your local HeartWare representative.
Quantity in commerce
9 units
Distribution
Worldwide Distribution - US Nationwide in the states of FL, MO, TN, KY, NY and the country of Germany.
Product codes and lots
Product Codes: ASY00116 and ASY00281 Serial # All patients with Drivelines splice repair performed prior April 2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1832-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1832-2017Class IOngoingVoluntary: Firm initiated