HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit
Heartware began a recall of HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been... on . FDA classified it as Class I and gives the reason as "In April 2015, HeartWare, now a part of Medtronic, notified users of a safety issue related to the HVAD Pump Driveline Splice Kit. Medtronic now intends to send a follow-up communication to identified clinicians...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1832-2017
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- HeartWare, Inc.
- Product
- HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been identified. These repairs are limited to the portion of the driveline that is outside the patient boundary (exit side).
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- In April 2015, HeartWare, now a part of Medtronic, notified users of a safety issue related to the HVAD Pump Driveline Splice Kit. Medtronic now intends to send a follow-up communication to identified clinicians informing them that a new, improved splice kit is available.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- HeartWare, now a part of Medtronic sent an Urgent Medical Device Recall letter on April 2015 and a follow-up letter on March 2017, to customers via overnight mail and/or email.. The letters identified the affected product, problem and actions to be taken. Customers were instructed to review the recall notice, forward the notice to individuals within their organization who need to be aware of the notice, complete, sign, and return the Acknowledgement Form . For questions contact your local HeartWare representative.
- Quantity in commerce
- 9 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of FL, MO, TN, KY, NY and the country of Germany.
- Product codes and lots
- Product Codes: ASY00116 and ASY00281 Serial # All patients with Drivelines splice repair performed prior April 2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1832-2017 | Class I | Ongoing | Voluntary: Firm initiated |