BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton...
Becton Dickinson & Company began a recall of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer... on . FDA classified it as Class II and gives the reason as "BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1746-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- BD sent an "Urgent Product Correction Letters and Customer Product Correction Response Forms" dated March 16, 2017 via email or UPS to their customers and distributors.
- Quantity in commerce
- 13,064,500 tubes (non-expired)
- Distribution
- Worldwide - US Nationwide Distribution
- Product codes and lots
- Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1746-2017 | Class II | Terminated | Voluntary: Firm initiated |