ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single...
Roho Inc. began a recall of ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology on . FDA classified it as Class II and gives the reason as "When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1921-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ROHO, Inc.
- Product
- ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.
- Product code
- KIC Cushion, Flotation
- Reason for recall
- When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- On 3/13/17, firm began notifying customers of recall via emailed or mailed letter. The notification sent to dealers and distributors includes a request to forward the communication to their customer and send an acknowledgement back once all of their customers have received a copy of the communication. Customers (product users) will not be required to send the product they are using back. Instead, the recall letter provided two options: 1) keep the cushion and follow additional set-up instructions provided within the letter that would mitigate the risk or 2) contact the supplier to have the cushion replaced. Product that had been in use and is replaced will be destroyed (evidence to be provided to ROHO, Inc. if destroyed by a customer). Unused product that is still in stock with the firm will be destroyed.
- Quantity in commerce
- 5646 units
- Distribution
- Nationwide, Sweden, Germany, New Zealand, Finland, Australia, Singapore, Korea, Luxembourg, Czech Republic, South Africa, Canada, Switzerland, United Arab Emirates, Spain, Austria, Hong Kong, Iceland, Japan, Canada, Italy, Ireland, France, United Kingdom
- Product codes and lots
- Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1921-2017 | Class II | Terminated | Voluntary: Firm initiated |