Merge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors...
Merge Healthcare, Inc. began a recall of Merge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors, measures, and records physiologic data from a human... on . FDA classified it as Class II and gives the reason as "Use of the software may show an incorrect value to the user when viewing the Fractional Flow Reserve (FFR) results during recording". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1778-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Use of the software may show an incorrect value to the user when viewing the Fractional Flow Reserve (FFR) results during recording.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- MERGE sent an Urgent Medical Device Recall letter dated April 4, 2016 to all affected customers via e-mail the week of April 4, 2016. If no email address was available, the letters were issued via certified mail. A second letter dated July 11, 2016 was issued to non-responders. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that all users of the affected product be provided with the notice. Customers were also asked to reply using the enclosed form and the return addressed envelope. Customers with questions were instructed to send an email to [email protected]. For questions regarding this recall call 877-741-5369.
- Quantity in commerce
- 24 sites potentially have the affected versions
- Distribution
- Nationwide Distribution including AZ, CA, CO, FL, GA, KS, LA, MI, MO, OH, OK, TN, TX, VA, WA, WI, and WV.
- Product codes and lots
- Versions 9.10.0, 9.20.0, 9.20.1, 9.20.2, 9.30.3, and 9.40.0.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1778-2017 | Class II | Terminated | Voluntary: Firm initiated |