Z-1772-2017Class II

Delta, Catalog Number: MS18597 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio...

Draeger Medical Systems, Inc. / initiated 2017-03-28 / Terminated

Draegar Medical Systems, Inc. began a recall of Delta, Catalog Number: MS18597 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Product Usage: The devices are intended... on . FDA classified it as Class II and gives the reason as "It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1772-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger Medical Systems, Inc.
Product
Delta, Catalog Number: MS18597 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
FDA root cause
Under Investigation by firm
Action
On March 28, 2017, US consignees were sent an Urgent Medical Recall letter and Customer Reply and Order card. Delta family monitors running software version 10.0 in facilities that also have at least one Scio module will be downgraded to software version VF9.1 as a temporary solution Free of Charge. Draeger Medical Systems, Inc. is developing a new software version to resolve the issue (VF10.1). Once available, all Delta family monitors that were running software version VF10.0 will be upgraded with VF10.1 Free of Charge. For all other countries, the responsible local Draeger Sales and Service will mail the letter and card.
Quantity in commerce
2,156 (total for all devices)
Distribution
Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
Product codes and lots
VF10.0 software

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1772-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1772-2017Class IITerminatedVoluntary: Firm initiated
Z-1773-2017Class IITerminatedVoluntary: Firm initiated
Z-1774-2017Class IITerminatedVoluntary: Firm initiated