Artis zee, Angiographic x-ray system
Siemens Medical Solutions USA, Inc began a recall of Artis zee, Angiographic x-ray system on . FDA classified it as Class II and gives the reason as "Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1824-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee, Angiographic x-ray system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next reactivation. Following reactivation, the system no longer initializes as intended.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens mailed a Customer Safety Advisory Notice (CSAN) AX017/17/S to affected customers on 03/28/2017 to inform them of the corrective action that will be performed to prevent a possible hazard to persons and equipment. The notice also explained the underlying issue, the impact on system operation, potential risk to patients and actions they are taking to correct the issue.
- Quantity in commerce
- 3 units distributed in U.S.
- Distribution
- Distributed to: VA, CA, WI
- Product codes and lots
- Device Model Numbers: 10280959, 10094137, 10094141, Serial numbers: 160449 147208, 153744
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1824-2017 | Class II | Terminated | Voluntary: Firm initiated |