Artis zee, Angiographic x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2017-03-28 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Artis zee, Angiographic x-ray system on . FDA classified it as Class II and gives the reason as "Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1824-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis zee, Angiographic x-ray system
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next reactivation. Following reactivation, the system no longer initializes as intended.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens mailed a Customer Safety Advisory Notice (CSAN) AX017/17/S to affected customers on 03/28/2017 to inform them of the corrective action that will be performed to prevent a possible hazard to persons and equipment. The notice also explained the underlying issue, the impact on system operation, potential risk to patients and actions they are taking to correct the issue.
Quantity in commerce
3 units distributed in U.S.
Distribution
Distributed to: VA, CA, WI
Product codes and lots
Device Model Numbers: 10280959, 10094137, 10094141, Serial numbers: 160449 147208, 153744

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1824-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1824-2017Class IITerminatedVoluntary: Firm initiated