Z-1872-2017Class II
PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile...
Cp Medical / initiated 2017-02-02 / Open, Classified
CP Medical Inc began a recall of PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes L923, L459, L493, L762, L463, L922... on . FDA classified it as Class II and gives the reason as "The tensile strength minimum as directed by USP <881> was not met over the length of the product's shelf life". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1872-2017
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Cp Medical
- Product
- PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes L923, L459, L493, L762, L463, L922, L495, L463, L497, L494, L493
- Product code
- GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
- Reason for recall
- The tensile strength minimum as directed by USP <881> was not met over the length of the product's shelf life.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- The affected distributors/customers were contacted directly via telephone or by e-mail on February 6-7, 2017. Those end users not contacted by CP Medical directly will be contacted by their distributor
- Quantity in commerce
- 504 boxes/12 sutures - 6,048 individual sutures
- Distribution
- AL, AZ, AR, CA, CO, CT, FL, ID, IL, IN, KY, LA, MT, NE, NJ, NM, NY, OK, OR, SC, TX, WA and South Africa
- Product codes and lots
- Lot Numbers - 150323-04, 150324-04, 150520-10, 150615-01, 150616-08, 150709-01, 151229-64, 160414-69, 160510-59, 160617-52, 160824-50
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1872-2017 | Class II | Ongoing | Voluntary: Firm initiated |