Z-1874-2017Class I

Newport Medical HT70 and HT70- Plus Ventilators, Rx Only

Newport Medical Instruments, Inc. / initiated 2017-03-30 / Terminated / root cause: software

Newport Medical Instruments Inc began a recall of Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or... on . FDA classified it as Class I and gives the reason as "Medtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1874-2017
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Newport Medical Instruments, Inc.
Product
Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs). The Newport HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The Newport HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Medtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Medtronic sent an Urgent Field Corrective Action Notice dated April 3, 2017, to all consignees via Federal Express or certified mail. The letter informed consignees of the remote potential for ventilator reset without accompanying alarm and the actions they should take. Customers were requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. Customers were also advised that a software service update will be available in May 2017 to address this issue. Medtronic issued a Press Release related to this voluntary field action on April 5, 2017. Consignees with questions were advised to contact the Technical Support Department immediately at 800-255-6774. For questions regarding this recall call 203-492-5415.
Quantity in commerce
12,966
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Canada, Australia, Brazil, Japan, Korea, Mexico, Bahrain, Egypt, Greece Hungary, Ireland, Israel, Jordan, Kazakhstan, Lithuania, Mauritius Namibia, Norway, Pakistan, Poland, Qatar, Romania, Russian Federation South Africa, Spain, Turkey, United Arab Emirates, United Kingdom Algeria, Armenia, Austria, Bahrain, Belarus, Belgium, Botswana Bulgaria, Cyprus, Czech Republic, Denmark, Egypt, Finland, France Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Kenya, Lebanon Lithuania, Macedonia, Malawi, Malta, Mauritius, Mozambique, Netherlands Norway, Oman, Poland, Portugal, Qatar, Republic of Moldova, Romania Saudi Arabia, Serbia, Slovakia, Sudan, Sweden, Switzerland, Tunisia Turkey, Uganda, Ukraine, United Kingdom, Yemen, and Zimbabwe
Product codes and lots
All Newport Medical HT70 and HT70- Plus Ventilators

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1874-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1874-2017Class ITerminatedVoluntary: Firm initiated