Z-1866-2017Class II

BD Viper LT software, version 3.00H

Becton Dickinson & Co / initiated 2015-12-14 / Terminated / root cause: software

Bd Diagnostic began a recall of BD Viper LT software, version 3.00H on . FDA classified it as Class II and gives the reason as "BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1866-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Viper LT software, version 3.00H
Product code
OOI Real Time Nucleic Acid Amplification System
Reason for recall
BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff threshold which has the potential to impact test results.
Root cause tag
software
FDA root cause
Software design
Action
Recall action was initiated on December 14, 2015 when software version 3.00H was removed from the two customer instruments. The 2 affected US customers were contacted via email and were also mailed a letter via UPS overnight express.
Quantity in commerce
2
Distribution
2 US customers, 0 Ex-US regions, and 0 distributors 0.
Product codes and lots
Catalog Number   443157 443153 443155 443156 442947 443902  Version Number    3.00H  3.00H

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1866-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1866-2017Class IITerminatedVoluntary: Firm initiated