da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000...
Intuitive Surgical, Inc. began a recall of da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic... on . FDA classified it as Class II and gives the reason as "Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1819-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic Surgery: The EndoWrist Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Firm sent e-mails on 3-17-17 to affected sites instructing them to STOP USE of the XI Suction Irrigator and RMA. and return all unused Xi Suction/irrigation instruments for credit. A formal letters was sent to affected sites on March 31, 2017. Letters advised of the reason for recall with details on product name and product number. Letters requested that any unused devices be returned to Intuitive via the standard RMA process. All personnel who use the Da Vinci system should be made aware of the issue. Response form should be completed online or by returning copy enclosed with the letter.
- Quantity in commerce
- 15 x 6 packs (90 units)
- Distribution
- US Only - one location each in AL, CO, KS, NV, NY, and 2 in TX
- Product codes and lots
- All
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1819-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY